The impact of Lenstar LS 900 optical biometry on intraocular lens power calculation accuracy: A prospective clinical study of 95 Iraqi patients with age-related cataract

Authors

  • Mahmood OS Department of Surgery, College of Medicine, Mustansiriyah University, Baghdad, Iraq
  • Al-Khafaji YARN Department of Surgery, College of Medicine, Mustansiriyah University, Baghdad, Iraq
  • Mohammad Ali FF Department of Surgery, College of Medicine, Kirkuk University, Kirkuk, Iraq

DOI:

https://doi.org/10.38029/babcockuniv.med.j..v9i3.1565

Keywords:

Intraocular lens calculation, Lenstar technology, Coherence reflectometry, Ocular biometry, Axial length measurement, Corneal curvature

Abstract

Objective: To evaluate the Lenstar LS 900 optical low-coherence reflectometry biometer for calculating intraocular lens (IOL) power in Iraqi patients having cataract surgery.

Methods: This prospective observational study recruited 95 adults (95 eyes) with age-related cataracts between July 2024 and April 2025 where in the method, was Preoperative biometry was performed on all participants with the Lenstar LS 900, so can which measured axial length, corneal curvature, anterior chamber depth, and lens thickness. The SRK/T, Haigis, and Hoffer Q formulas were used to calculate IOL power. A single experienced surgeon carried out phacoemulsification. The main outcome was the mean absolute error (MAE) of predicted and achieved refraction at one-month follow-up. Also found in the section results were the secondary outcomes, which were calculation time, proportion of eyes within the range of ±0.50 dioptres (D) of target refraction, and patient satisfaction.

Results: The value of mean MAE was 0.24 ± 0.11 D. Lenstar predictions were accurate (within ±0.50 D of target) in 92.6% (88/95) of eyes, in addition to found value of Mean IOL calculation time was 5.2 +/- 1.1 minutes. Most commonly used formula was the SRK/T formula (52.6%), also the section of Axial length was normal (22.0–24.0 mm) in 52.6% of eyes, while on the other hand, found Eighty-five-point three percent of patients were satisfied with their visual results postoperatively. There were no severe complications.

Conclusion: Lenstar LS 900 optical biometry offers high predictive accuracy and efficient workflow in IOL power calculation in patients with age-related cataract.

Downloads

Published

2026-09-30

How to Cite

Mahmood, O., Al-Khafaji, Y., & Mohammad, F. (2026). The impact of Lenstar LS 900 optical biometry on intraocular lens power calculation accuracy: A prospective clinical study of 95 Iraqi patients with age-related cataract. Babcock University Medical Journal, 9(3), 348–354. https://doi.org/10.38029/babcockuniv.med.j.v9i3.1565